Pressure Support 4 – Expiratory Cycling

This is likely the most important of the four tutorials on Pressure Support Ventilation. As you may recall, PS is an unusual mode of ventilation because it is flow cycled – that is – the ventilator cycles to expiration as specific, user set, percentage of peak flow. The default expiratory sensitivity is usually around 25%. Expiratory dys-synchrony is frequently missed by bedside clinicians who have not been schooled in waveform analysis. This tutorial covers everything you need to know. @ccmtutorials http://www.ccmtutorials.org

Next time I am going to commence a series of tutorials on hypoxia-hypoxemia. This will start with a discussion about how we measure hypoxemia – in particular oxyhemoglobin saturation (Tutorial 12). I will then go on to discuss atelectasis, shunt, ventilation-perfusion mismatch and introduce oxygen therapy (Tutorial 13).